Previous image
ChraLab CL100 Lab Stainless steel drying oven 100L
Next image
Zoom image

ChraLab CL100 Lab Stainless steel drying oven 100L

  • ChraLab CL100 oven with high level of quality and reliability to meet various requirements for drying, sterilization and heating applications.
  • This technically masterpiece is made of high quality stainless steel - easy to clean - automatic ventilation and control system, super heat protection and precision controlled heating technology. 
  • Precise drying, heating, ageing, burn-in and hardening in research, science, industry and quality assurance.
  • Stainless steel drying oven
  • Intuitive and easy-to-use operating menu
  • Air exchange rates and air flap position electronically controllable
  • Pre-heated fresh air prevents temperature fluctuations
  • Almost exclusive use of high-quality, corrosion-resistant and easily cleanable stainless steel for the working chamber and housing
  • $0.00

Precise drying, heating, ageing, burn-in and hardening in research, science, industry and quality assurance.


The universally applicable lab oven ChraLab CL100 classic appliance for temperature control in science, research and material tests in industry. The technologically perfected masterpiece made of high-quality, hygienic, easy-to-clean stainless steel leaves nothing to be desired in terms of ventilation and control technology, over temperature protection and precisely tuned heating technology.

Dimensions
Dimension_UF30Defined and programme-controlled fan speed
Air exchange rates and air flap position can be controlled electronically at the Control. More inlet and outlet openings lead to a higher air exchange and reduced drying times. Various applications recommend or even require controlled ventilation. When drying powder, sand or corn, reducing the ventilation prevents undesired swirls

Other applications like testing of wires or cables demand for defined air exchange rates. Appliances feature easy programming of temperature and air exchange rates with the Control software.

Fresh air is preheated
Temperature deviations caused by fresh air can influence sample characteristics or prolong drying. In ChraLab CL200 ovens, the fresh air is therefore fed through a pre-heating chamber and introduced into the working chamber.Fresh_Air

Manufacturer Specs

ChraLab CL100 Specifications
Temperature
Set temperature range in °Cmin. 5°C above ambient up to +300°C
Temperature1 Pt100 sensor DIN class A in 4-wire-circuit
Control technology
Language settingGerman, English, Spanish, French, Polish, Czech, Hungarian
ControlAdaptive multifunctional digital PID-microprocessor controller with high-definition colour display
TimerDigital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAITthe process time does not start until the set temperature is reached
Calibrationthree freely selectable temperature values
adjustable parameterstemperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime
Ventilation
Convectionnatural convection
Fresh air admixtureadjustment of pre-heated fresh air admixture by air flap control in 10 % steps
Ventvent connection with restrictor flap
Communication
Documentationprogramme stored in case of power failure
ProgrammingCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with CONTROL software available as accessory (on demand).
Safety
Temperature controladjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Autodiagnostic systemfor fault analysis
Standard equipment
Doorfully insulated stainless steel door with 2-point locking (compression door lock)
Internals2 stainless steel grid(s), electropolished
Works calibration certificateCalibration at +160°C
Stainless steel interior
Interioreasy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume100 l
Dimensionsw(A) x h(B) x d(C): 550 x 450 x 400 mm 
Max. number of internals9
Max. loading of chamber300 kg
Max. loading per internal20 kg
Textured stainless steel casing
Dimensionsw(D) x h(E) x d(F): 740 x 860 x 580 mm 
Housingrear zinc-plated steel
Electrical data
Voltage Electrical load230 V, 50/60 Hz
approx. 3400 W
Voltage Electrical load115 V, 50/60 Hz
approx. 1800 W
Ambient conditions
Set UpThe distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Altitude of installationmax. 2,000 m above sea level
Ambient temperature+5 °C to +40 °C
Humidity rhmax. 80 %, non-condensing
Overvoltage categoryII
Pollution degree2
Packing/shipping data
Transport informationThe appliances must be transported upright

Dimensions approx incl. cartonw x h x d: 900 x 1300 x 900 cm
Net weightapprox. 100 kg
Gross weight cartonapprox. 160 kg
    • SKU
      ChraLab CL100